Title: Stability of Spectrometric Characteristics of CsI:Tl Detectors Depending on the Surface Treatment Method
Other Titles: Стабильность спектрометрических характеристик кристаллов CsI:Tl в зависимости от способа обработки поверхности
Authors: Vydai, Yury T.
Tarasov, Vladimir A.
Kudin, Alexander M.
Andryushchenko, Lubov A.
Ananenko, Andrey A.
Keywords: CsI:Tl crystals
Stability of characteristics
Surface treatment
alpha-particles
Issue Date: 2006
Publisher: Pleiades Publishing, Inc.,
Citation: Instruments and Experimental Techniques, Vol. 49, 3, pp. 314–317
Abstract: The influence of various mechanochemical methods used to treat the entrance surface of a CsI:Tl scintillator on the specific light output and the pulse-height resolution during detection of α particles with energies of 2 or 5 MeV is described. Nonlinearity between the light output and the energy of α particles increases from 7% for a polished surface to 35% after grinding, which leads to a degraded pulse-height resolution. A method for treating the CsI:Tl surface is proposed that minimizes the light-output nonlinearity and ensures stability of spectrometric characteristics in time.
URI: http://repositsc.nuczu.edu.ua/handle/123456789/5348